| ECG PERFORMANCE |
| ECG Leads |
Simultaneous 12 channel ECG acquisition |
| Recording Channels |
3CH+3RHY, 3CH+1RHY, 6CH+1RHY, 12CH, 1CH+3, Cabrera report, 1CH long time (1 min, 3 min, 5 min, 10 min, 20 min, 30 min), special beat report (text, guide, vector) |
| Gain |
2.5, 5, 10, 20, auto (I–aVF: 10, V1–V6: 5) mm/mV |
| Print Speed |
5, 12.5, 25, 50, 100 mm/s |
| Sampling Rate |
Digital sampling rate: 8,000 Hz |
| Filters |
AC (50/60 Hz, -20 dB or better) Muscle (25–35 Hz, -3 dB or better) Bionet baseline drift (0.05 Hz, 0.1 Hz, 0.2 Hz, -3 dB or better) Low pass (off, 40 Hz, 100 Hz, 150 Hz) |
| Measurements |
Heart rate (30–300 bpm, ±3 bpm), PR/RR interval, QRS duration, QT/QTc interval, P–R–T axis, SV1/RV5/R+S amplitude |
| Interpretation |
Bionet ECG analysis algorithm, University of Glasgow ECG analysis algorithm |
| DISPLAY AND INTERFACE |
| Display |
8" color TFT wide display (1024 × 600), 12-channel preview ECG waveform |
| User Interface |
Capacitive touch screen (alphanumeric and symbols available), function keys |
| PATIENT INFORMATION |
| Patient Data |
ID, name, date of birth, age, gender, height, weight, race, smoke, department, room no., study description, accession no., referring physician |
| RECORDER AND PAPER |
| Printer |
Thermal print head, Z-fold paper |
| Report Paper |
A4 210 mm (8.3") × 300 mm (11.8"), Letter 215 mm (8.5") × 280 mm (11") |
| Paper Size |
Half A4 210 mm × 150 mm, Half Letter 215 mm × 140 mm |
| Print Resolution |
Vertical: 8 dot/mm Horizontal: 16 dot/mm (0.125 mm pitch) |
| ELECTRICAL AND SIGNAL |
| Internal Noise |
20 uV (p-p) max |
| Input Impedance |
≥ 50 MΩ |
| Input Voltage Range |
±5 mV |
| CMRR |
> 105 dB |
| DC Offset |
≥ ±400 mV |
| Frequency Response |
0.05–200 Hz (within -3 dB) |
| Protection |
Isolated, defibrillation and ESU protected |
| DATA, STORAGE AND CONNECTIVITY |
| Signal Quality |
Pacemaker pulse detection, lead fault detection, signal saturation detection |
| Data Storage |
Internal storage for 500 data, built-in memory |
| Communication |
LAN, WiFi (option), USB flash drive, USB barcode scanner |
| POWER AND BATTERY |
| Line Power |
Input: 100–240 VAC, 50/60 Hz, 1.5–0.75 A Output: 15 VDC, 4.2 A |
| Battery Type |
Replaceable and rechargeable lithium ion, 10.8 V, 6500 mA |
| Battery Capacity |
10 hours normal use, or print 350 ECGs (12 channel at 25 mm/s and 10 mm/mV) or spiro pages, recharge to full in 3 hours (device turned off) |
| SAFETY AND COMPLIANCE |
| Safety Conformity |
Class I, Type CF applied parts (ECG electrodes), Type B applied parts (spirometer handle) |
| ENVIRONMENT |
| Operating Temperature |
10–40°C |
| Operating Humidity |
30–85% |
| Operating Pressure |
700–1060 hPa |
| Storage Temperature |
-20–60°C |
| Storage Humidity |
10–95% |
| Storage Pressure |
500–1060 hPa |
| PHYSICAL |
| Dimensions |
Main body: 286 mm (W) × 350 mm (D) × 140 mm (H) |
| Weight |
Approximately 4.5 kg (max) |